A wearable bluetooth LE sensor for patient monitoring during MRI scans.

Christian Vogt, Jonas Reber, Daniel Waltisberg, Lars Buthe, Josip Marjanovic, Niko Munzenrieder, Klaas P Pruessmann, Gerhard Troster

PAPERpubmedAnnual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference2016engineeringEffect: mixedEvidence: Low

Abstract

This paper presents a working prototype of a wearable patient monitoring device capable of recording the heart rate, blood oxygen saturation, surface temperature and humidity during an magnetic resonance imaging (MRI) experiment. The measured values are transmitted via Bluetooth low energy (LE) and displayed in real time on a smartphone on the outside of the MRI room. During 7 MRI image acquisitions of at least 1 min and a total duration of 25 min no Bluetooth data packets were lost. The raw measurements of the light intensity for the photoplethysmogram based heart rate measurement shows an increased noise floor by 50LSB (least significant bit) during the MRI operation, whereas the temperature and humidity readings are unaffected. The device itself creates a magnetic resonance (MR) signal loss with a radius of 14 mm around the device surface and shows no significant increase in image noise of an acquired MRI image due to its radio frequency activity. This enables continuous and unobtrusive patient monitoring during MRI scans.

AI evidence extraction

At a glance
Study type
engineering
Effect direction
mixed
Population
—
Sample size
—
Exposure
RF bluetooth LE wearable device during MRI · 7 MRI image acquisitions of at least 1 min; total duration 25 min
Evidence strength
Low
Confidence: 74% · Peer-reviewed: yes

Main findings

In 7 MRI image acquisitions (total 25 min), no Bluetooth data packets were lost. Photoplethysmogram light-intensity raw measurements showed an increased noise floor by 50 LSB during MRI operation, while temperature and humidity readings were unaffected. The device caused MR signal loss with a 14 mm radius around its surface and showed no significant increase in MRI image noise attributable to its RF activity.

Outcomes measured

  • Bluetooth data packet loss during MRI
  • Noise floor of photoplethysmogram (light intensity) during MRI operation
  • Temperature and humidity sensor readings during MRI
  • MRI signal loss/artifact radius around device
  • MRI image noise due to device radio frequency activity

Limitations

  • No human participant population described in the abstract
  • Sample size for subjects not reported
  • Exposure parameters (e.g.
  • Bluetooth frequency
  • RF power/SAR
  • MRI field strength) not reported
  • Short total test duration (25 min)
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": "RF",
    "source": "bluetooth LE wearable device during MRI",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": "7 MRI image acquisitions of at least 1 min; total duration 25 min"
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "Bluetooth data packet loss during MRI",
    "Noise floor of photoplethysmogram (light intensity) during MRI operation",
    "Temperature and humidity sensor readings during MRI",
    "MRI signal loss/artifact radius around device",
    "MRI image noise due to device radio frequency activity"
  ],
  "main_findings": "In 7 MRI image acquisitions (total 25 min), no Bluetooth data packets were lost. Photoplethysmogram light-intensity raw measurements showed an increased noise floor by 50 LSB during MRI operation, while temperature and humidity readings were unaffected. The device caused MR signal loss with a 14 mm radius around its surface and showed no significant increase in MRI image noise attributable to its RF activity.",
  "effect_direction": "mixed",
  "limitations": [
    "No human participant population described in the abstract",
    "Sample size for subjects not reported",
    "Exposure parameters (e.g., Bluetooth frequency, RF power/SAR, MRI field strength) not reported",
    "Short total test duration (25 min)"
  ],
  "evidence_strength": "low",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "Bluetooth low energy",
    "wearable sensor",
    "patient monitoring",
    "MRI compatibility",
    "photoplethysmogram",
    "heart rate",
    "SpO2",
    "temperature",
    "humidity",
    "RF activity",
    "MRI artifact"
  ],
  "suggested_hubs": [
    {
      "slug": "medical-mri",
      "weight": 0.9,
      "reason": "Wearable Bluetooth LE device tested for operation and imaging artifacts during MRI scans."
    }
  ]
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 2885

A wearable bluetooth LE sensor for patient monitoring during MRI scans. · Non-native EMF