Variation of radio frequency induced power deposition due to second surrounding tissue.

Mikhail Kozlov, Gregor Schaefers

PAPERpubmedAnnual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference2016engineeringEffect: unclearEvidence: Insufficient

Abstract

We investigated RF-induced power deposition (p) generated by insulated leads with different lengths located in proximity to two tissues. The first tissue surrounding the leads was shaped as a cylinder with a diameter of 11 mm and located in a very large box filled with the second tissue. Lead electromagnetic models and p substantially depended on electrical properties of both tissues. If the cylinder and the box were filled with the same tissue, discrepancy to results of the two-tissue setup was up to 280%.

AI evidence extraction

At a glance
Study type
engineering
Effect direction
unclear
Population
—
Sample size
—
Exposure
RF
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes

Main findings

RF-induced power deposition generated by insulated leads depended substantially on the electrical properties of two surrounding tissues. When both surrounding regions were modeled as the same tissue (single-tissue setup), results differed from the two-tissue setup by up to 280%.

Outcomes measured

  • RF-induced power deposition (p) near insulated leads
  • Dependence of lead electromagnetic models on surrounding tissue electrical properties

Limitations

  • Frequency and exposure parameters were not reported in the abstract.
  • No biological/clinical outcomes were described
  • focus was on modeling/power deposition.
  • Sample size and whether results are experimental or purely computational are not specified in the abstract.
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": "RF",
    "source": "other",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "RF-induced power deposition (p) near insulated leads",
    "Dependence of lead electromagnetic models on surrounding tissue electrical properties"
  ],
  "main_findings": "RF-induced power deposition generated by insulated leads depended substantially on the electrical properties of two surrounding tissues. When both surrounding regions were modeled as the same tissue (single-tissue setup), results differed from the two-tissue setup by up to 280%.",
  "effect_direction": "unclear",
  "limitations": [
    "Frequency and exposure parameters were not reported in the abstract.",
    "No biological/clinical outcomes were described; focus was on modeling/power deposition.",
    "Sample size and whether results are experimental or purely computational are not specified in the abstract."
  ],
  "evidence_strength": "insufficient",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "radiofrequency",
    "power deposition",
    "insulated leads",
    "tissue electrical properties",
    "two-tissue model",
    "electromagnetic modeling"
  ],
  "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 2886

Variation of radio frequency induced power deposition due to second surrounding tissue. · Non-native EMF