Ultrafast volumetric B1 (+) mapping for improved radiofrequency shimming in 3 tesla body MRI.
Abstract
PURPOSE: To evaluate the use of the recently proposed ultrafast B1 (+) mapping approach DREAM (Dual Refocusing Echo Acquisition Mode) for a refinement of patient adaptive radiofrequency (RF) shimming. MATERIALS AND METHODS: Volumetric DREAM B1 (+) calibration scans centered in the upper abdomen were acquired in 20 patients and three volunteers with written informed consent at a clinical dual source 3 Tesla (T) MR system. Based on these data, RF transmit settings were optimized by central-slice based RF-shimming (CS-RF shim) and by a refined, multi-slice adaptive approach (MS-RF shim). Simulations were performed to compare flip angle accuracy and B1 (+) homogeneity (cv = stddev/mean) achieved by CS-RF shim versus MS-RF shim for transversal and coronal slices, and for volume shimming on the spine. RESULTS: By MS-RF shim, mean deviation from nominal flip angle was reduced to less than 11% in all slices, all targets, and all subjects. Relative improvements in B1 (+) cv (MS-RF shim versus CS-RF) were up to 14%/39%/47% in transversal slices/coronal slices/ spine area. CONCLUSION: Volumetric information about B1 (+) can be used to further improve the accuracy and homogeneity of the B1 (+) field yielding higher diagnostic confidence, and will also be of value for various quantitative methods which are sensitive to flip angle imperfections.
AI evidence extraction
Main findings
Using multi-slice adaptive RF shimming (MS-RF shim) based on volumetric DREAM B1+ calibration scans, mean deviation from nominal flip angle was reduced to less than 11% across slices/targets/subjects. Relative improvements in B1+ homogeneity (cv) for MS-RF shim versus central-slice RF shimming were reported up to 14% (transversal), 39% (coronal), and 47% (spine area).
Outcomes measured
- B1+ mapping performance (DREAM)
- Flip angle accuracy (deviation from nominal)
- B1+ homogeneity (coefficient of variation
- cv)
- Performance of RF shimming approaches (CS-RF shim vs MS-RF shim)
View raw extracted JSON
{
"study_type": "engineering",
"exposure": {
"band": "RF",
"source": "MRI",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": "20 patients and three volunteers",
"sample_size": 23,
"outcomes": [
"B1+ mapping performance (DREAM)",
"Flip angle accuracy (deviation from nominal)",
"B1+ homogeneity (coefficient of variation, cv)",
"Performance of RF shimming approaches (CS-RF shim vs MS-RF shim)"
],
"main_findings": "Using multi-slice adaptive RF shimming (MS-RF shim) based on volumetric DREAM B1+ calibration scans, mean deviation from nominal flip angle was reduced to less than 11% across slices/targets/subjects. Relative improvements in B1+ homogeneity (cv) for MS-RF shim versus central-slice RF shimming were reported up to 14% (transversal), 39% (coronal), and 47% (spine area).",
"effect_direction": "benefit",
"limitations": [],
"evidence_strength": "low",
"confidence": 0.74,
"peer_reviewed_likely": "yes",
"keywords": [
"B1+ mapping",
"DREAM",
"Dual Refocusing Echo Acquisition Mode",
"radiofrequency shimming",
"RF transmit settings",
"3 Tesla",
"body MRI",
"flip angle",
"B1+ homogeneity",
"upper abdomen"
],
"suggested_hubs": [
{
"slug": "mri-rf",
"weight": 0.9,
"reason": "Study evaluates RF (B1+) mapping and RF shimming methods in a 3T clinical MRI system."
}
]
}AI can be wrong. Always verify against the paper.