Intensified microwave coupled by a loop antenna (diameter of 3 mm) has been employed to enhance the laser-induced breakdown spectroscopy (LIBS) emission. In this method, a laser plasma was induced on Gd₂O₃ sample at a reduced pressure by focusing a pulsed Nd:YAG laser (532 nm, 10 ns, 5 mJ) at a local point, at which electromagnetic field was produced by introducing microwave radiation using loop antenna. The plasma emission was significantly enhanced by absorbing the microwave radiation, resulting in high-temperature plasma and long-lifetime plasma emission. By using this method, the enhancement of Gd lines was up to 32 times, depending upon the emission lines observed. A linear calibration curve of Ca contained in the Gd₂O₃ sample was made. The detection limit of Ca was approximately 2 mg/kg. This present method is very useful for identification of trace elements in nuclear fuel and radioactive materials.

AI evidence extraction

At a glance
Study type
engineering
Effect direction
benefit
Population
—
Sample size
—
Exposure
microwave
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes

Main findings

Microwave radiation introduced via a 3 mm diameter loop antenna was absorbed by the laser-induced plasma and significantly enhanced LIBS emission, producing higher-temperature and longer-lifetime plasma emission. Reported enhancement of Gd spectral lines was up to 32×, and the Ca detection limit in Gd2O3 was approximately 2 mg/kg.

Outcomes measured

  • LIBS plasma emission intensity
  • Plasma temperature (high-temperature plasma)
  • Plasma emission lifetime
  • Enhancement factor of Gd spectral lines
  • Calibration curve for Ca in Gd2O3
  • Detection limit for Ca (~2 mg/kg)

Limitations

  • Microwave frequency and exposure parameters (e.g.
  • power) not stated in the abstract
  • No sample size or number of measurements reported
  • Study is an instrumentation/analytical method description rather than a biological/health assessment
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": "microwave",
    "source": "antenna-coupled microwave (loop antenna)",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "LIBS plasma emission intensity",
    "Plasma temperature (high-temperature plasma)",
    "Plasma emission lifetime",
    "Enhancement factor of Gd spectral lines",
    "Calibration curve for Ca in Gd2O3",
    "Detection limit for Ca (~2 mg/kg)"
  ],
  "main_findings": "Microwave radiation introduced via a 3 mm diameter loop antenna was absorbed by the laser-induced plasma and significantly enhanced LIBS emission, producing higher-temperature and longer-lifetime plasma emission. Reported enhancement of Gd spectral lines was up to 32×, and the Ca detection limit in Gd2O3 was approximately 2 mg/kg.",
  "effect_direction": "benefit",
  "limitations": [
    "Microwave frequency and exposure parameters (e.g., power) not stated in the abstract",
    "No sample size or number of measurements reported",
    "Study is an instrumentation/analytical method description rather than a biological/health assessment"
  ],
  "evidence_strength": "insufficient",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "microwave",
    "loop antenna",
    "laser-induced breakdown spectroscopy",
    "LIBS",
    "Nd:YAG laser",
    "plasma emission",
    "Gd2O3",
    "trace element detection",
    "calibration curve",
    "detection limit"
  ],
  "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 3110

Enhancement of LIBS emission using antenna-coupled microwave. · Non-native EMF