  "id": 3110,
  "url": "https://nnemf.org/research/3110-enhancement-of-libs-emission-using-antenna-coupled-microwave",
  "title": "Enhancement of LIBS emission using antenna-coupled microwave.",
  "authors": [
    "Ali Khumaeni",
    "Tampo Motonobu",
    "Akaoka Katsuaki",
    "Miyabe Masabumi",
    "Wakaida Ikuo"
  ],
  "journal": "Optics express",
  "year": "2013",
  "doi": "10.1364/oe.21.029755",
  "pmid": "24514526",
  "abstract": "Intensified microwave coupled by a loop antenna (diameter of 3 mm) has been employed to enhance the laser-induced breakdown spectroscopy (LIBS) emission. In this method, a laser plasma was induced on Gd₂O₃ sample at a reduced pressure by focusing a pulsed Nd:YAG laser (532 nm, 10 ns, 5 mJ) at a local point, at which electromagnetic field was produced by introducing microwave radiation using loop antenna. The plasma emission was significantly enhanced by absorbing the microwave radiation, resulting in high-temperature plasma and long-lifetime plasma emission. By using this method, the enhancement of Gd lines was up to 32 times, depending upon the emission lines observed. A linear calibration curve of Ca contained in the Gd₂O₃ sample was made. The detection limit of Ca was approximately 2 mg/kg. This present method is very useful for identification of trace elements in nuclear fuel and radioactive materials.",
  "effect": "benefit",
  "evidence": "insufficient",
  "extraction": {
    "studyType": "engineering",
    "studyTypeLabel": "engineering",
    "effectDirection": "benefit",
    "evidenceLabel": "Insufficient",
    "population": "—",
    "sampleSize": "—",
    "exposure": "microwave",
    "confidencePct": "74",
    "peerReviewed": "yes",
    "mainFindings": "Microwave radiation introduced via a 3 mm diameter loop antenna was absorbed by the laser-induced plasma and significantly enhanced LIBS emission, producing higher-temperature and longer-lifetime plasma emission. Reported enhancement of Gd spectral lines was up to 32×, and the Ca detection limit in Gd2O3 was approximately 2 mg/kg.",
    "outcomes": [
      "LIBS plasma emission intensity",
      "Plasma temperature (high-temperature plasma)",
      "Plasma emission lifetime",
      "Enhancement factor of Gd spectral lines",
      "Calibration curve for Ca in Gd2O3",
      "Detection limit for Ca (~2 mg/kg)"
    ],
    "limitations": [
      "Microwave frequency and exposure parameters (e.g.",
      "power) not stated in the abstract",
      "No sample size or number of measurements reported",
      "Study is an instrumentation/analytical method description rather than a biological/health assessment"
    ],
    "keywords": [
      "microwave",
      "loop antenna",
      "laser-induced breakdown spectroscopy",
      "LIBS",
      "Nd:YAG laser",
      "plasma emission",
      "Gd2O3",
      "trace element detection",
      "calibration curve",
      "detection limit"
    ],
    "hubs": [],
    "raw": "{\n  \"study_type\": \"engineering\",\n  \"exposure\": {\n    \"band\": \"microwave\",\n    \"source\": \"antenna-coupled microwave (loop antenna)\",\n    \"frequency_mhz\": null,\n    \"sar_wkg\": null,\n    \"duration\": null\n  },\n  \"population\": null,\n  \"sample_size\": null,\n  \"outcomes\": [\n    \"LIBS plasma emission intensity\",\n    \"Plasma temperature (high-temperature plasma)\",\n    \"Plasma emission lifetime\",\n    \"Enhancement factor of Gd spectral lines\",\n    \"Calibration curve for Ca in Gd2O3\",\n    \"Detection limit for Ca (~2 mg/kg)\"\n  ],\n  \"main_findings\": \"Microwave radiation introduced via a 3 mm diameter loop antenna was absorbed by the laser-induced plasma and significantly enhanced LIBS emission, producing higher-temperature and longer-lifetime plasma emission. Reported enhancement of Gd spectral lines was up to 32×, and the Ca detection limit in Gd2O3 was approximately 2 mg/kg.\",\n  \"effect_direction\": \"benefit\",\n  \"limitations\": [\n    \"Microwave frequency and exposure parameters (e.g., power) not stated in the abstract\",\n    \"No sample size or number of measurements reported\",\n    \"Study is an instrumentation/analytical method description rather than a biological/health assessment\"\n  ],\n  \"evidence_strength\": \"insufficient\",\n  \"confidence\": 0.74,\n  \"peer_reviewed_likely\": \"yes\",\n  \"keywords\": [\n    \"microwave\",\n    \"loop antenna\",\n    \"laser-induced breakdown spectroscopy\",\n    \"LIBS\",\n    \"Nd:YAG laser\",\n    \"plasma emission\",\n    \"Gd2O3\",\n    \"trace element detection\",\n    \"calibration curve\",\n    \"detection limit\"\n  ],\n  \"suggested_hubs\": []\n}"
  },
  "synthesis": null,
  "synthesisStatus": "pending",
  "originalHref": "https://pubmed.ncbi.nlm.nih.gov/24514526/",
  "sourceUrl": "https://rfsafe.org/mel/paper.php?id=3110",
  "updated": "2026-02-13 07:06:37"
}