Abstract

This paper describes wireless measurement platform, which uses inductive power and data transmission and provides up to four analog measurement channels. Platform is powered by inductive coupling using the 125 kHz magnetic field. Data transmission is realized by modulating the magnetic field. The platform has up to four 10-bit analog channels. The sample rate of the platform is 237 samples per second, which is divided between the used channels as needed. With 2.85 V supply voltage the measurement platform has power consumption less than 1.5 mW. Nevertheless the end application's total power consumption depends on the measurement set-up. Except for the antenna coil, the measurement platform is made using the commercially available components. Main target application for the platform is implantable electrocardiogram (ECG) measurement. At this stage ECG measurements are done with one channel. Additional channels are available for example for other ECG channels, temperature or acceleration measurements.

AI evidence extraction

At a glance
Study type
engineering
Effect direction
unclear
Population
—
Sample size
—
Exposure
ELF inductive power/data transmission (magnetic field) · 0.125 MHz
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes

Main findings

A wireless measurement platform using inductive coupling with a 125 kHz magnetic field is described, providing up to four 10-bit analog channels and a total sample rate of 237 samples/s. With a 2.85 V supply, the platform power consumption is reported as <1.5 mW; ECG measurements were performed with one channel at this stage.

Outcomes measured

  • Wireless measurement platform performance (analog channels
  • sample rate
  • power consumption)
  • Inductive power and data transmission using 125 kHz magnetic field
  • Target application: implantable ECG measurement

Limitations

  • No biological or health outcomes are reported
  • this is a device/platform description.
  • No exposure levels (e.g.
  • field strength) or dosimetry (e.g.
  • SAR) are provided.
  • No study population or sample size is described.

Suggested hubs

  • occupational-exposure (0.7)
    Mentions magnetic-field-based inductive power/data transmission, but no occupational setting is described.
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": "ELF",
    "source": "inductive power/data transmission (magnetic field)",
    "frequency_mhz": 0.125,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "Wireless measurement platform performance (analog channels, sample rate, power consumption)",
    "Inductive power and data transmission using 125 kHz magnetic field",
    "Target application: implantable ECG measurement"
  ],
  "main_findings": "A wireless measurement platform using inductive coupling with a 125 kHz magnetic field is described, providing up to four 10-bit analog channels and a total sample rate of 237 samples/s. With a 2.85 V supply, the platform power consumption is reported as <1.5 mW; ECG measurements were performed with one channel at this stage.",
  "effect_direction": "unclear",
  "limitations": [
    "No biological or health outcomes are reported; this is a device/platform description.",
    "No exposure levels (e.g., field strength) or dosimetry (e.g., SAR) are provided.",
    "No study population or sample size is described."
  ],
  "evidence_strength": "insufficient",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "inductive coupling",
    "magnetic field",
    "125 kHz",
    "wireless power",
    "data transmission",
    "implantable",
    "ECG",
    "analog measurement channels"
  ],
  "suggested_hubs": [
    {
      "slug": "occupational-exposure",
      "weight": 0.1,
      "reason": "Mentions magnetic-field-based inductive power/data transmission, but no occupational setting is described."
    }
  ]
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 3477

Wireless platform for multi-channel analog measurements. · Non-native EMF