Analysis of current density and specific absorption rate in biological tissue surrounding an air-core type of transcutaneous transformer for an artificial heart.
Abstract
This paper reports on the specific absorption rate (SAR) and the current density analysis of biological tissue surrounding an air-core type of transcutaneous transformer for an artificial heart. The electromagnetic field in the biological tissue surrounding the transformer was analyzed by the transmission-line modeling method, and the SAR and current density as a function of frequency (200k-1 MHz) for a transcutaneous transmission of 20 W were calculated. The model's biological tissue has three layers including the skin, fat and muscle. As a result, the SAR in the vicinity of the transformer is sufficiently small and the normalized SAR value, which is divided by the ICNIRP's basic restriction, is 7 x 10(-3) or less. On the contrary, the current density is slightly in excess of the ICNIRP's basic restrictions as the frequency falls and the output voltage rises. Normalized current density is from 0.2 to 1.2. In addition, the layer in which the current's density is maximized depends on the frequency, the muscle in the low frequency (<700 kHz) and the skin in the high frequency (>700 kHz). The result shows that precision analysis taking into account the biological properties is very important for developing the transcutaneous transformer for TAH.
AI evidence extraction
Main findings
Using transmission-line modeling for a 20 W transcutaneous transformer (200 kHz–1 MHz) with a 3-layer tissue model (skin, fat, muscle), SAR near the transformer was reported as sufficiently small; normalized SAR (relative to ICNIRP basic restriction) was ≤7×10^-3. In contrast, current density slightly exceeded ICNIRP basic restrictions at lower frequencies and higher output voltages, with normalized current density ranging from 0.2 to 1.2; the tissue layer with maximum current density depended on frequency (muscle <700 kHz; skin >700 kHz).
Outcomes measured
- specific absorption rate (SAR)
- current density
- comparison to ICNIRP basic restrictions
Limitations
- Modeling/simulation study (transmission-line modeling) rather than measurements in humans or animals
- No sample size or empirical exposure measurements reported in abstract
- Specific SAR and current density absolute values not provided in abstract (only normalized values/ranges)
Suggested hubs
- who-icnirp (0.7)Results are explicitly normalized to and compared against ICNIRP basic restrictions.
- occupational-exposure (0.7)Involves device-related EMF exposure assessment, though not explicitly occupational.
View raw extracted JSON
{
"study_type": "engineering",
"exposure": {
"band": "ELF",
"source": "transcutaneous transformer (artificial heart)",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": null,
"sample_size": null,
"outcomes": [
"specific absorption rate (SAR)",
"current density",
"comparison to ICNIRP basic restrictions"
],
"main_findings": "Using transmission-line modeling for a 20 W transcutaneous transformer (200 kHz–1 MHz) with a 3-layer tissue model (skin, fat, muscle), SAR near the transformer was reported as sufficiently small; normalized SAR (relative to ICNIRP basic restriction) was ≤7×10^-3. In contrast, current density slightly exceeded ICNIRP basic restrictions at lower frequencies and higher output voltages, with normalized current density ranging from 0.2 to 1.2; the tissue layer with maximum current density depended on frequency (muscle <700 kHz; skin >700 kHz).",
"effect_direction": "mixed",
"limitations": [
"Modeling/simulation study (transmission-line modeling) rather than measurements in humans or animals",
"No sample size or empirical exposure measurements reported in abstract",
"Specific SAR and current density absolute values not provided in abstract (only normalized values/ranges)"
],
"evidence_strength": "low",
"confidence": 0.78,
"peer_reviewed_likely": "yes",
"keywords": [
"transcutaneous transformer",
"artificial heart",
"TAH",
"specific absorption rate",
"SAR",
"current density",
"ICNIRP",
"transmission-line modeling",
"200 kHz",
"1 MHz",
"skin",
"fat",
"muscle"
],
"suggested_hubs": [
{
"slug": "who-icnirp",
"weight": 0.8,
"reason": "Results are explicitly normalized to and compared against ICNIRP basic restrictions."
},
{
"slug": "occupational-exposure",
"weight": 0.2,
"reason": "Involves device-related EMF exposure assessment, though not explicitly occupational."
}
]
}AI can be wrong. Always verify against the paper.