Determination of blood longitudinal relaxation time (T1) at high magnetic field strengths.
Abstract
In this study, a circulation system was used to measure T(1) values of bovine blood under physiological conditions at field strengths of 4.7, 7 and 9.4 T. Results show that T(1) increases linearly with magnetic field B(0) and can be described with the equation T(1)=129 ms/T B(0)+1167 ms for magnetic field strengths between 1.5 and 9.4 T.
AI evidence extraction
At a glance
Study type
Other
Effect direction
unclear
Population
Bovine blood (circulating system)
Sample size
—
Exposure
MRI (static magnetic field B0)
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes
Main findings
Using a circulation system under physiological conditions, bovine blood T1 increased linearly with magnetic field strength (B0) across 4.7, 7, and 9.4 T. The relationship was described as T1 = 129 ms/T * B0 + 1167 ms for magnetic field strengths between 1.5 and 9.4 T.
Outcomes measured
- Blood longitudinal relaxation time (T1)
Limitations
- Sample size not reported in abstract
- Study uses bovine blood in a circulation system
- generalizability to humans not stated
- No health outcomes assessed
- outcome is a measurement parameter (T1)
View raw extracted JSON
{
"study_type": "other",
"exposure": {
"band": null,
"source": "MRI (static magnetic field B0)",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": "Bovine blood (circulating system)",
"sample_size": null,
"outcomes": [
"Blood longitudinal relaxation time (T1)"
],
"main_findings": "Using a circulation system under physiological conditions, bovine blood T1 increased linearly with magnetic field strength (B0) across 4.7, 7, and 9.4 T. The relationship was described as T1 = 129 ms/T * B0 + 1167 ms for magnetic field strengths between 1.5 and 9.4 T.",
"effect_direction": "unclear",
"limitations": [
"Sample size not reported in abstract",
"Study uses bovine blood in a circulation system; generalizability to humans not stated",
"No health outcomes assessed; outcome is a measurement parameter (T1)"
],
"evidence_strength": "insufficient",
"confidence": 0.74,
"peer_reviewed_likely": "yes",
"keywords": [
"MRI",
"static magnetic field",
"B0",
"T1 relaxation time",
"blood",
"high field strength",
"4.7 T",
"7 T",
"9.4 T"
],
"suggested_hubs": []
}AI can be wrong. Always verify against the paper.
AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.
Comments
Log in to comment.
No comments yet.
NNEMF synthesis
Synthesis pending.