Performance of a radio link between a base station and a medical implant utilising the MICS standard.

Anders J Johansson

PAPERpubmedConference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference2004engineeringEffect: unclearEvidence: Insufficient

Abstract

Modern medical implants are of increasing complexity and with that, the need for fast and flexible communication with them grows. A wireless system is preferable and an inductive link is the most commonly used. But it has the drawback of a very short range, essentially limited to having the external transceiver touching the patient. The Medical Implant Communication System, MICS, is a standard aimed at improving the communication distance. It operates at a higher frequency band between 402 MHz and 405 MHz. We have by simulations and measurements investigated the channel properties of this band and calculated the link performance for a typical implant application. The result is a link speed between a base station and a bedridden patient of 600 kbit bits per second with a bit error rate of 2% in the downlink to the implant and 1 % in the uplink to the base station. Conclusions on the necessary complexity of the base station are also given.

AI evidence extraction

At a glance
Study type
engineering
Effect direction
unclear
Population
—
Sample size
—
Exposure
RF 100–1000 MHz
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: unknown

Main findings

Using simulations and measurements of the MICS band (402–405 MHz), the authors calculated link performance for a typical implant application. They report a link speed of 600 kbit/s between a base station and a bedridden patient, with bit error rates of 2% (downlink to implant) and 1% (uplink to base station), and provide conclusions on required base-station complexity.

Outcomes measured

  • Channel properties at 402–405 MHz (MICS band)
  • Link speed (kbps)
  • Bit error rate (downlink/uplink)
  • Base station complexity requirements

Limitations

  • No health outcomes assessed (focus is link/channel performance).
  • Sample size and measurement setup details are not provided in the abstract.
  • Exposure metrics such as SAR or field strength are not reported in the abstract.
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": "RF",
    "source": "medical implant communication (MICS) base station/implant link",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "Channel properties at 402–405 MHz (MICS band)",
    "Link speed (kbps)",
    "Bit error rate (downlink/uplink)",
    "Base station complexity requirements"
  ],
  "main_findings": "Using simulations and measurements of the MICS band (402–405 MHz), the authors calculated link performance for a typical implant application. They report a link speed of 600 kbit/s between a base station and a bedridden patient, with bit error rates of 2% (downlink to implant) and 1% (uplink to base station), and provide conclusions on required base-station complexity.",
  "effect_direction": "unclear",
  "limitations": [
    "No health outcomes assessed (focus is link/channel performance).",
    "Sample size and measurement setup details are not provided in the abstract.",
    "Exposure metrics such as SAR or field strength are not reported in the abstract."
  ],
  "evidence_strength": "insufficient",
  "confidence": 0.74,
  "peer_reviewed_likely": "unknown",
  "keywords": [
    "medical implant communication system",
    "MICS",
    "402–405 MHz",
    "radio link",
    "base station",
    "implant",
    "channel properties",
    "bit error rate",
    "link speed"
  ],
  "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 3612

Performance of a radio link between a base station and a medical implant utilising the MICS standard. · Non-native EMF