  "id": 3612,
  "url": "https://nnemf.org/research/3612-performance-of-a-radio-link-between-a-base",
  "title": "Performance of a radio link between a base station and a medical implant utilising the MICS standard.",
  "authors": [
    "Anders J Johansson"
  ],
  "journal": "Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference",
  "year": "2004",
  "doi": "10.1109/iembs.2004.1403620",
  "pmid": "17272140",
  "abstract": "Modern medical implants are of increasing complexity and with that, the need for fast and flexible communication with them grows. A wireless system is preferable and an inductive link is the most commonly used. But it has the drawback of a very short range, essentially limited to having the external transceiver touching the patient. The Medical Implant Communication System, MICS, is a standard aimed at improving the communication distance. It operates at a higher frequency band between 402 MHz and 405 MHz. We have by simulations and measurements investigated the channel properties of this band and calculated the link performance for a typical implant application. The result is a link speed between a base station and a bedridden patient of 600 kbit bits per second with a bit error rate of 2% in the downlink to the implant and 1 % in the uplink to the base station. Conclusions on the necessary complexity of the base station are also given.",
  "effect": "unclear",
  "evidence": "insufficient",
  "extraction": {
    "studyType": "engineering",
    "studyTypeLabel": "engineering",
    "effectDirection": "unclear",
    "evidenceLabel": "Insufficient",
    "population": "—",
    "sampleSize": "—",
    "exposure": "RF 100–1000 MHz",
    "confidencePct": "74",
    "peerReviewed": "unknown",
    "mainFindings": "Using simulations and measurements of the MICS band (402–405 MHz), the authors calculated link performance for a typical implant application. They report a link speed of 600 kbit/s between a base station and a bedridden patient, with bit error rates of 2% (downlink to implant) and 1% (uplink to base station), and provide conclusions on required base-station complexity.",
    "outcomes": [
      "Channel properties at 402–405 MHz (MICS band)",
      "Link speed (kbps)",
      "Bit error rate (downlink/uplink)",
      "Base station complexity requirements"
    ],
    "limitations": [
      "No health outcomes assessed (focus is link/channel performance).",
      "Sample size and measurement setup details are not provided in the abstract.",
      "Exposure metrics such as SAR or field strength are not reported in the abstract."
    ],
    "keywords": [
      "medical implant communication system",
      "MICS",
      "402–405 MHz",
      "radio link",
      "base station",
      "implant",
      "channel properties",
      "bit error rate",
      "link speed"
    ],
    "hubs": [],
    "raw": "{\n  \"study_type\": \"engineering\",\n  \"exposure\": {\n    \"band\": \"RF\",\n    \"source\": \"medical implant communication (MICS) base station/implant link\",\n    \"frequency_mhz\": null,\n    \"sar_wkg\": null,\n    \"duration\": null\n  },\n  \"population\": null,\n  \"sample_size\": null,\n  \"outcomes\": [\n    \"Channel properties at 402–405 MHz (MICS band)\",\n    \"Link speed (kbps)\",\n    \"Bit error rate (downlink/uplink)\",\n    \"Base station complexity requirements\"\n  ],\n  \"main_findings\": \"Using simulations and measurements of the MICS band (402–405 MHz), the authors calculated link performance for a typical implant application. They report a link speed of 600 kbit/s between a base station and a bedridden patient, with bit error rates of 2% (downlink to implant) and 1% (uplink to base station), and provide conclusions on required base-station complexity.\",\n  \"effect_direction\": \"unclear\",\n  \"limitations\": [\n    \"No health outcomes assessed (focus is link/channel performance).\",\n    \"Sample size and measurement setup details are not provided in the abstract.\",\n    \"Exposure metrics such as SAR or field strength are not reported in the abstract.\"\n  ],\n  \"evidence_strength\": \"insufficient\",\n  \"confidence\": 0.74,\n  \"peer_reviewed_likely\": \"unknown\",\n  \"keywords\": [\n    \"medical implant communication system\",\n    \"MICS\",\n    \"402–405 MHz\",\n    \"radio link\",\n    \"base station\",\n    \"implant\",\n    \"channel properties\",\n    \"bit error rate\",\n    \"link speed\"\n  ],\n  \"suggested_hubs\": []\n}"
  },
  "synthesis": null,
  "synthesisStatus": "pending",
  "originalHref": "https://pubmed.ncbi.nlm.nih.gov/17272140/",
  "sourceUrl": "https://rfsafe.org/mel/paper.php?id=3612",
  "updated": "2026-02-13 07:30:52"
}