  "id": 6537,
  "url": "https://nnemf.org/research/6537-investigation-of-non-ionizing-radiation-hazards-from-physiotherapy",
  "title": "[Investigation of non-ionizing radiation hazards from physiotherapy equipment in 16 medical institutions].",
  "authors": [
    "Jia-xi He",
    "Wei Zhou",
    "Hai-li Qiu",
    "Guang-tao Yang"
  ],
  "journal": "Zhonghua lao dong wei sheng zhi ye bing za zhi = Zhonghua laodong weisheng zhiyebing zazhi = Chinese journal of industrial hygiene and occupational diseases",
  "year": "2013",
  "doi": null,
  "pmid": "24370361",
  "abstract": "OBJECTIVE: To investigate the non-ionizing radiation hazards from physiotherapy equipment in medical institutions and to explore feasible control measures for occupational diseases. METHODS: On-site measurement and assessment of ultra-high-frequency radiation, high-frequency electromagnetic field, microwave radiation, and laser radiation were carried out in 16 medical institutions using the methods in the Measurement of Physical Agents in Workplace (GBZ/T189-2007). RESULTS: All the investigated medical institutions failed to take effective protective measures against non-ionizing radiation. Of the 17 ultra-short wave therapy apparatus, 70.6%, 47.1%, and 17.64% had a safe intensity of ultra-high-frequency radiation on the head, chest, and abdomen, respectively. Of the 4 external high-frequency thermotherapy apparatus, 100%, 75%, and 75%had a safe intensity of high-frequency electromagnetic field on the head, chest, and abdomen, respectively. In addition, the intensities of microwave radiation and laser radiation produced by the 18 microwave therapy apparatus and 12 laser therapeutic apparatus met national health standards. CONCLUSION: There are non-ionizing radiation hazards from physiotherapy equipment in medical institutions, and effective prevention and control measures are necessary.",
  "effect": "mixed",
  "evidence": "low",
  "extraction": {
    "studyType": "exposure_assessment",
    "studyTypeLabel": "exposure assessment",
    "effectDirection": "mixed",
    "evidenceLabel": "Low",
    "population": "Workers in medical institutions (occupational setting)",
    "sampleSize": "16",
    "exposure": "microwave",
    "confidencePct": "74",
    "peerReviewed": "yes",
    "mainFindings": "On-site measurements in 16 medical institutions found that none had effective protective measures against non-ionizing radiation. For 17 ultra-short wave therapy apparatus, the proportion with safe ultra-high-frequency radiation intensity was 70.6% (head), 47.1% (chest), and 17.64% (abdomen). For 4 external high-frequency thermotherapy apparatus, the proportion with safe high-frequency electromagnetic field intensity was 100% (head), 75% (chest), and 75% (abdomen). Intensities from 18 microwave therapy apparatus and 12 laser therapeutic apparatus met national health standards.",
    "outcomes": [
      "Workplace intensity measurements of ultra-high-frequency radiation",
      "Workplace intensity measurements of high-frequency electromagnetic fields",
      "Workplace intensity measurements of microwave radiation",
      "Workplace intensity measurements of laser radiation",
      "Compliance with national health standards / safe intensity at head",
      "chest",
      "abdomen",
      "Presence/absence of protective measures"
    ],
    "limitations": [
      "No specific frequencies",
      "measurement values",
      "or exposure limits are reported in the abstract",
      "Health outcomes are not assessed",
      "only exposure levels and compliance are described",
      "Sampling/selection of institutions and devices is not described in the abstract"
    ],
    "keywords": [
      "non-ionizing radiation",
      "occupational exposure",
      "physiotherapy equipment",
      "ultra-high-frequency radiation",
      "high-frequency electromagnetic field",
      "microwave therapy",
      "laser therapy",
      "workplace measurement",
      "China"
    ],
    "hubs": [
      {
        "slug": "occupational-exposure",
        "weight": "0.7",
        "reason": "Workplace measurements of non-ionizing radiation from physiotherapy equipment and discussion of occupational disease prevention/control."
      }
    ],
    "raw": "{\n  \"study_type\": \"exposure_assessment\",\n  \"exposure\": {\n    \"band\": null,\n    \"source\": \"physiotherapy equipment in medical institutions (ultra-short wave therapy, external high-frequency thermotherapy, microwave therapy, laser therapeutic apparatus)\",\n    \"frequency_mhz\": null,\n    \"sar_wkg\": null,\n    \"duration\": null\n  },\n  \"population\": \"Workers in medical institutions (occupational setting)\",\n  \"sample_size\": 16,\n  \"outcomes\": [\n    \"Workplace intensity measurements of ultra-high-frequency radiation\",\n    \"Workplace intensity measurements of high-frequency electromagnetic fields\",\n    \"Workplace intensity measurements of microwave radiation\",\n    \"Workplace intensity measurements of laser radiation\",\n    \"Compliance with national health standards / safe intensity at head, chest, abdomen\",\n    \"Presence/absence of protective measures\"\n  ],\n  \"main_findings\": \"On-site measurements in 16 medical institutions found that none had effective protective measures against non-ionizing radiation. For 17 ultra-short wave therapy apparatus, the proportion with safe ultra-high-frequency radiation intensity was 70.6% (head), 47.1% (chest), and 17.64% (abdomen). For 4 external high-frequency thermotherapy apparatus, the proportion with safe high-frequency electromagnetic field intensity was 100% (head), 75% (chest), and 75% (abdomen). Intensities from 18 microwave therapy apparatus and 12 laser therapeutic apparatus met national health standards.\",\n  \"effect_direction\": \"mixed\",\n  \"limitations\": [\n    \"No specific frequencies, measurement values, or exposure limits are reported in the abstract\",\n    \"Health outcomes are not assessed; only exposure levels and compliance are described\",\n    \"Sampling/selection of institutions and devices is not described in the abstract\"\n  ],\n  \"evidence_strength\": \"low\",\n  \"confidence\": 0.74,\n  \"peer_reviewed_likely\": \"yes\",\n  \"keywords\": [\n    \"non-ionizing radiation\",\n    \"occupational exposure\",\n    \"physiotherapy equipment\",\n    \"ultra-high-frequency radiation\",\n    \"high-frequency electromagnetic field\",\n    \"microwave therapy\",\n    \"laser therapy\",\n    \"workplace measurement\",\n    \"China\"\n  ],\n  \"suggested_hubs\": [\n    {\n      \"slug\": \"occupational-exposure\",\n      \"weight\": 0.9,\n      \"reason\": \"Workplace measurements of non-ionizing radiation from physiotherapy equipment and discussion of occupational disease prevention/control.\"\n    }\n  ]\n}"
  },
  "synthesis": null,
  "synthesisStatus": "pending",
  "originalHref": "https://pubmed.ncbi.nlm.nih.gov/24370361/",
  "sourceUrl": "https://rfsafe.org/mel/paper.php?id=6537",
  "updated": "2026-02-16 02:04:08"
}