Patients with pacemakers or defibrillators do not need to worry about e-Cars: An observational study
Abstract
Patients with pacemakers or defibrillators do not need to worry about e-Cars: An observational study
AI evidence extraction
At a glance
Study type
cross sectional
Effect direction
no_effect
Population
Patients with cardiac implantable electronic devices (CIED), including pacemakers or defibrillators
Sample size
108
Exposure
electric cars (driving and charging) · roller-bench test-driving; charging (high current charging mentioned)
Evidence strength
Low
Confidence: 78% · Peer-reviewed: yes
Main findings
In 108 patients with CIEDs exposed to electromagnetic fields from four common electric cars during roller-bench test-driving and charging, no abnormalities in CIED function or programming were observed by ECG recordings and device interrogation. The largest magnetic field detected was 116.5 μT along the charging cable during high-current charging; cabin field strength was 2.1–3.6 μT.
Outcomes measured
- Electromagnetic interference (EMI) effects on CIED function/programming (e.g.
- oversensing with pacing-inhibition
- inappropriate therapy
- mode-switching)
- Measured magnetic field strength during driving/charging (μT)
Limitations
- Frequency characteristics of the electromagnetic fields were not reported in the abstract.
- Exposure duration and detailed exposure protocol (time
- distances/positions) were not fully specified in the abstract.
- Study design described as prospective
- cross-sectional
- no control/comparator group is described in the abstract.
- Only four electric car models were tested
- generalizability to other vehicles/charging systems is not established in the abstract.
Suggested hubs
- occupational-exposure (0.7)Includes exposure during driving/charging scenarios, though not explicitly occupational.
View raw extracted JSON
{
"study_type": "cross_sectional",
"exposure": {
"band": null,
"source": "electric cars (driving and charging)",
"frequency_mhz": null,
"sar_wkg": null,
"duration": "roller-bench test-driving; charging (high current charging mentioned)"
},
"population": "Patients with cardiac implantable electronic devices (CIED), including pacemakers or defibrillators",
"sample_size": 108,
"outcomes": [
"Electromagnetic interference (EMI) effects on CIED function/programming (e.g., oversensing with pacing-inhibition, inappropriate therapy, mode-switching)",
"Measured magnetic field strength during driving/charging (μT)"
],
"main_findings": "In 108 patients with CIEDs exposed to electromagnetic fields from four common electric cars during roller-bench test-driving and charging, no abnormalities in CIED function or programming were observed by ECG recordings and device interrogation. The largest magnetic field detected was 116.5 μT along the charging cable during high-current charging; cabin field strength was 2.1–3.6 μT.",
"effect_direction": "no_effect",
"limitations": [
"Frequency characteristics of the electromagnetic fields were not reported in the abstract.",
"Exposure duration and detailed exposure protocol (time, distances/positions) were not fully specified in the abstract.",
"Study design described as prospective, cross-sectional; no control/comparator group is described in the abstract.",
"Only four electric car models were tested; generalizability to other vehicles/charging systems is not established in the abstract."
],
"evidence_strength": "low",
"confidence": 0.78,
"peer_reviewed_likely": "yes",
"keywords": [
"electric cars",
"electromagnetic interference",
"EMI",
"cardiac implantable electronic devices",
"CIED",
"pacemaker",
"defibrillator",
"charging cable",
"magnetic field",
"microtesla"
],
"suggested_hubs": [
{
"slug": "medical-implants",
"weight": 0.95,
"reason": "Assesses electromagnetic interference risk to pacemakers/defibrillators (CIEDs) from electric cars."
},
{
"slug": "occupational-exposure",
"weight": 0.25,
"reason": "Includes exposure during driving/charging scenarios, though not explicitly occupational."
}
]
}AI can be wrong. Always verify against the paper.
AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.