High-Fidelity 3D Stray Magnetic Field Mapping of Smartphones to Address Safety Considerations with Active Implantable Electronic Medical

PAPERmanual2023exposure assessmentEffect: harmEvidence: Low

Abstract

High-Fidelity 3D Stray Magnetic Field Mapping of Smartphones to Address Safety Considerations with Active Implantable Electronic Medical Devices Saha N, Millward JM, Herrmann CJJ, Rahimi F, Han H, Lacour P, Blaschke F, Niendorf T. High-Fidelity 3D Stray Magnetic Field Mapping of Smartphones to Address Safety Considerations with Active Implantable Electronic Medical Devices. Sensors (Basel). 2023 Jan 20;23(3):1209. doi: 10.3390/s23031209. Abstract Case reports indicate that magnets in smartphones could be a source of electromagnetic interference (EMI) for active implantable medical devices (AIMD), which could lead to device malfunction, compromising patient safety. Recognizing this challenge, we implemented a high-fidelity 3D magnetic field mapping (spatial resolution 1 mm) setup using a three-axis Hall probe and teslameter, controlled by a robot (COSI Measure). With this setup, we examined the stray magnetic field of an iPhone 13 Pro, iPhone 12, and MagSafe charger to identify sources of magnetic fields for the accurate risk assessment of potential interferences with AIMDs. Our measurements revealed that the stray fields of the annular array of magnets, the wide-angle camera, and the speaker of the smartphones exceeded the 1 mT limit defined by ISO 14117:2019. Our data-driven safety recommendation is that an iPhone 13 Pro should be kept at least 25 mm away from an AIMD to protect it from unwanted EMI interactions. Our study addresses safety concerns due to potential device-device interactions between smartphones and AIMDs and will help to define data-driven safety guidelines. We encourage vendors of electronic consumer products (ECP) to provide information on the magnetic fields of their products and advocate for the inclusion of smartphones in the risk assessment of EMI with AIMDs. : mdpi.com

AI evidence extraction

At a glance
Study type
exposure assessment
Effect direction
harm
Population
—
Sample size
—
Exposure
smartphone
Evidence strength
Low
Confidence: 78% · Peer-reviewed: yes

Main findings

Using high-fidelity 3D magnetic field mapping (1 mm spatial resolution), the stray fields from the annular magnet array, wide-angle camera, and speaker of tested smartphones exceeded the 1 mT limit defined by ISO 14117:2019. The authors recommend keeping an iPhone 13 Pro at least 25 mm away from an AIMD to reduce risk of unwanted EMI interactions.

Outcomes measured

  • 3D stray magnetic field mapping of smartphones and MagSafe charger
  • Magnetic field exceedance of 1 mT limit (ISO 14117:2019) relevant to electromagnetic interference risk for active implantable medical devices (AIMDs)
  • Recommended minimum separation distance between iPhone 13 Pro and AIMD (25 mm)

Suggested hubs

  • who-icnirp (0.7)
    Mentions an international standard (ISO 14117:2019) related to EMI limits for implantable devices; no direct WHO/ICNIRP content stated.
  • occupational-exposure (0.7)
    Mapped from the finder screen onto an existing hub.
View raw extracted JSON
{
  "study_type": "exposure_assessment",
  "exposure": {
    "band": null,
    "source": "smartphone",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "3D stray magnetic field mapping of smartphones and MagSafe charger",
    "Magnetic field exceedance of 1 mT limit (ISO 14117:2019) relevant to electromagnetic interference risk for active implantable medical devices (AIMDs)",
    "Recommended minimum separation distance between iPhone 13 Pro and AIMD (25 mm)"
  ],
  "main_findings": "Using high-fidelity 3D magnetic field mapping (1 mm spatial resolution), the stray fields from the annular magnet array, wide-angle camera, and speaker of tested smartphones exceeded the 1 mT limit defined by ISO 14117:2019. The authors recommend keeping an iPhone 13 Pro at least 25 mm away from an AIMD to reduce risk of unwanted EMI interactions.",
  "effect_direction": "harm",
  "limitations": [],
  "evidence_strength": "low",
  "confidence": 0.78,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "smartphones",
    "iPhone 13 Pro",
    "iPhone 12",
    "MagSafe charger",
    "stray magnetic field",
    "3D mapping",
    "electromagnetic interference",
    "EMI",
    "active implantable medical devices",
    "AIMD",
    "ISO 14117:2019",
    "Hall probe",
    "teslameter"
  ],
  "suggested_hubs": [
    {
      "slug": "who-icnirp",
      "weight": 0.2,
      "reason": "Mentions an international standard (ISO 14117:2019) related to EMI limits for implantable devices; no direct WHO/ICNIRP content stated."
    },
    {
      "slug": "occupational-exposure",
      "weight": 0,
      "reason": null
    }
  ]
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 575

High-Fidelity 3D Stray Magnetic Field Mapping of Smartphones to Address Safety Considerations with Active Implantable Electronic Medical · Non-native EMF