Extremely Low Frequency Electromagnetic Field from Convective Air Warming System on Temperature Selection and Distance.

Kwang Rae Cho, Myoung-Hun Kim, Myoung Jin Ko, Jae Wook Jung, Ki Hwa Lee, Yei-Heum Park, Yong Han Kim, Ki Hoon Kim, Jin Soo Kim

PAPERpubmedIranian journal of public health2014exposure assessmentEffect: unclearEvidence: Insufficient

Abstract

BACKGROUND: Hypothermia generates potentially severe complications in operating or recovery room. Forced air warmer is effective to maintain body temperature. Extremely low frequency electromagnetic field (ELF-EMF) is harmful to human body and mainly produced by electronic equipment including convective air warming system. We investigated ELF-EMF from convective air warming device on various temperature selection and distance for guideline to protect medical personnel and patients. METHODS: The intensity of ELF-EMF was measured as two-second interval for five minutes on various distance (0.1, 0.2, 0.3, 0.5 and 1meter) and temperature selection (high, medium, low and ambient). All of electrical devices were off including lamp, computer and air conditioner. Groups were compared using one-way ANOVA. P<0.05 was considered significant. RESULTS: Mean values of ELF-EMF on the distance of 30 cm were 18.63, 18.44, 18.23 and 17.92 milligauss (mG) respectively (high, medium, low and ambient temperature set). ELF-EMF of high temperature set was higher than data of medium, low and ambient set in all the distances. CONCLUSION: ELF-EMF from convective air warming system is higher in condition of more close location and higher temperature. ELF-EMF within thirty centimeters exceeds 2mG recommended by Swedish TCO guideline.

AI evidence extraction

At a glance
Study type
exposure assessment
Effect direction
unclear
Population
—
Sample size
—
Exposure
ELF convective air warming system (forced air warmer) · measured at 2-second intervals for 5 minutes at each condition
Evidence strength
Insufficient
Confidence: 78% · Peer-reviewed: yes

Main findings

ELF-EMF intensity decreased with increasing distance and was higher at higher temperature settings. At 30 cm, mean ELF-EMF values were 18.63, 18.44, 18.23, and 17.92 mG for high, medium, low, and ambient settings, respectively. The authors report that within 30 cm the measured ELF-EMF exceeded the 2 mG level recommended by the Swedish TCO guideline.

Outcomes measured

  • ELF-EMF intensity (milligauss
  • mG) by distance and temperature setting
  • Comparison to Swedish TCO 2 mG guideline threshold

Limitations

  • No frequency characteristics of the ELF-EMF reported
  • No information on device model/brand or measurement instrument/calibration reported in abstract
  • No direct health outcomes assessed
  • study measures field intensity only
  • Sample size/number of repeated measurements per condition not stated beyond timing description

Suggested hubs

  • occupational-exposure (0.7)
    Exposure measurements relevant to medical personnel working near operating/recovery room equipment.
View raw extracted JSON
{
  "study_type": "exposure_assessment",
  "exposure": {
    "band": "ELF",
    "source": "convective air warming system (forced air warmer)",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": "measured at 2-second intervals for 5 minutes at each condition"
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "ELF-EMF intensity (milligauss, mG) by distance and temperature setting",
    "Comparison to Swedish TCO 2 mG guideline threshold"
  ],
  "main_findings": "ELF-EMF intensity decreased with increasing distance and was higher at higher temperature settings. At 30 cm, mean ELF-EMF values were 18.63, 18.44, 18.23, and 17.92 mG for high, medium, low, and ambient settings, respectively. The authors report that within 30 cm the measured ELF-EMF exceeded the 2 mG level recommended by the Swedish TCO guideline.",
  "effect_direction": "unclear",
  "limitations": [
    "No frequency characteristics of the ELF-EMF reported",
    "No information on device model/brand or measurement instrument/calibration reported in abstract",
    "No direct health outcomes assessed; study measures field intensity only",
    "Sample size/number of repeated measurements per condition not stated beyond timing description"
  ],
  "evidence_strength": "insufficient",
  "confidence": 0.78,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "extremely low frequency",
    "ELF-EMF",
    "forced air warmer",
    "convective air warming system",
    "operating room",
    "exposure measurement",
    "milligauss",
    "distance",
    "temperature setting",
    "Swedish TCO guideline"
  ],
  "suggested_hubs": [
    {
      "slug": "occupational-exposure",
      "weight": 0.62,
      "reason": "Exposure measurements relevant to medical personnel working near operating/recovery room equipment."
    }
  ]
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 3013

Extremely Low Frequency Electromagnetic Field from Convective Air Warming System on Temperature Selection and Distance. · Non-native EMF