Liver lesion conspicuity during real-time MR-guided radiofrequency applicator placement using spoiled gradient echo and balanced steady-state free precession imaging.

Hansjörg Rempp, Henning Loh, Rüdiger Hoffmann, Eva Rothgang, Li Pan, Claus D Claussen, Stephan Clasen

PAPERpubmedJournal of magnetic resonance imaging : JMRI2014OtherEffect: unclearEvidence: Low

Abstract

PURPOSE: To retrospectively evaluate the conspicuity of liver lesions in a fluoroscopic spoiled gradient echo (GRE) and a balanced steady-state free precession (SSFP) magnetic resonance imaging (MRI) sequence. MATERIALS AND METHODS: In all, 103 patients with hepatocellular carcinomas (HCC) (41) or liver metastases (67) were treated using MR-guided radiofrequency ablation in a wide-bore 1.5 T scanner. A multislice real-time spoiled GRE sequence allowing for a T1 weighting (T1W) and a balanced SSFP sequence allowing for a T2/T1W contrast were used for MR guidance. The contrast-to-noise-ratio (CNR) of the lesions was calculated and lesion conspicuity was assessed retrospectively (easily detectable / difficult to detect / not detectable). RESULTS: HCC was easily detectable in 33/52% (GRE/SSFP), difficult to detect in 30/18%, and not detectable in 37/30% of the cases. Mean CNR varied widely (9.1 for GRE vs. 16.4 for SSFP). Liver metastases were easily detectable in 58/41% (GRE/SSFP), difficult to detect in 14/21%, and not detectable in 28/38% of the cases. Mean CNR for liver metastases was 11.5 (GRE) vs. 12.7 (SSFP). Twenty percent of all lesions could not be detected with either of the MR fluoroscopy sequences. CONCLUSION: MR fluoroscopy using GRE and SSFP contrast enabled real-time detectability of 80% of the liver lesions.

AI evidence extraction

At a glance
Study type
Other
Effect direction
unclear
Population
Patients with hepatocellular carcinoma (HCC) or liver metastases undergoing MR-guided radiofrequency ablation in a 1.5 T MRI scanner
Sample size
103
Exposure
RF
Evidence strength
Low
Confidence: 74% · Peer-reviewed: yes

Main findings

In 103 patients treated with MR-guided radiofrequency ablation, lesion detectability differed by sequence (spoiled GRE vs balanced SSFP) and lesion type. Overall, 20% of lesions could not be detected with either MR fluoroscopy sequence, implying 80% real-time detectability across sequences.

Outcomes measured

  • Liver lesion conspicuity (easily detectable / difficult to detect / not detectable)
  • Contrast-to-noise ratio (CNR)

Limitations

  • Retrospective evaluation
View raw extracted JSON
{
  "study_type": "other",
  "exposure": {
    "band": null,
    "source": null,
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": "Patients with hepatocellular carcinoma (HCC) or liver metastases undergoing MR-guided radiofrequency ablation in a 1.5 T MRI scanner",
  "sample_size": 103,
  "outcomes": [
    "Liver lesion conspicuity (easily detectable / difficult to detect / not detectable)",
    "Contrast-to-noise ratio (CNR)"
  ],
  "main_findings": "In 103 patients treated with MR-guided radiofrequency ablation, lesion detectability differed by sequence (spoiled GRE vs balanced SSFP) and lesion type. Overall, 20% of lesions could not be detected with either MR fluoroscopy sequence, implying 80% real-time detectability across sequences.",
  "effect_direction": "unclear",
  "limitations": [
    "Retrospective evaluation"
  ],
  "evidence_strength": "low",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "MRI",
    "MR fluoroscopy",
    "spoiled gradient echo",
    "balanced SSFP",
    "liver lesions",
    "hepatocellular carcinoma",
    "liver metastases",
    "MR-guided radiofrequency ablation",
    "contrast-to-noise ratio",
    "lesion conspicuity"
  ],
  "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.