Share

A mm-sized wirelessly powered and remotely controlled locomotive implant.

Daniel Pivonka, Anatoly Yakovlev, Ada S Y Poon, Teresa Meng

PAPERpubmedIEEE transactions on biomedical circuits and systems2012engineeringEffect: unclearEvidence: Insufficient

Abstract

A wirelessly powered and controlled implantable device capable of locomotion in a fluid medium is presented. Two scalable low-power propulsion methods are described that achieve roughly an order of magnitude better performance than existing methods in terms of thrust conversion efficiency. The wireless prototype occupies 0.6 mm × 1 mm in 65 nm CMOS with an external 2 mm × 2 mm receive antenna. The IC consists of a matching network, a rectifier, a bandgap reference, a regulator, a demodulator, a digital controller, and high-current drivers that interface directly with the propulsion system. It receives 500 μW from a 2 W 1.86 GHz power signal at a distance of 5 cm. Asynchronous pulse-width modulation on the carrier allows for data rates from 2.5-25 Mbps with energy efficiency of 0.5 pJ/b at 10 Mbps. The received data configures the propulsion system drivers, which are capable of driving up to 2 mA at 0.2 V and can achieve speed of 0.53 cm/sec in a 0.06 T magnetic field.

AI evidence extraction

At a glance
Study type
engineering
Effect direction
unclear
Population
—
Sample size
—
Exposure
RF wireless power signal (implantable device) · 1860 MHz
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes

Main findings

A mm-scale implantable locomotive device was wirelessly powered and controlled. The prototype received 500 µW from a 2 W, 1.86 GHz power signal at 5 cm distance and supported 2.5–25 Mbps data rates via asynchronous PWM on the carrier; the propulsion drivers enabled locomotion up to 0.53 cm/s in a 0.06 T magnetic field.

Outcomes measured

  • Wireless power transfer performance (received power at distance)
  • Implant device locomotion/propulsion performance (speed
  • thrust conversion efficiency)
  • Wireless data communication performance (data rate
  • energy efficiency)

Limitations

  • No biological/health outcomes are reported in the abstract.
  • No exposure metrics such as SAR or tissue heating are reported.
  • Study context appears to be device engineering
  • population/sample details are not provided.
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": "RF",
    "source": "wireless power signal (implantable device)",
    "frequency_mhz": 1860,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "Wireless power transfer performance (received power at distance)",
    "Implant device locomotion/propulsion performance (speed, thrust conversion efficiency)",
    "Wireless data communication performance (data rate, energy efficiency)"
  ],
  "main_findings": "A mm-scale implantable locomotive device was wirelessly powered and controlled. The prototype received 500 µW from a 2 W, 1.86 GHz power signal at 5 cm distance and supported 2.5–25 Mbps data rates via asynchronous PWM on the carrier; the propulsion drivers enabled locomotion up to 0.53 cm/s in a 0.06 T magnetic field.",
  "effect_direction": "unclear",
  "limitations": [
    "No biological/health outcomes are reported in the abstract.",
    "No exposure metrics such as SAR or tissue heating are reported.",
    "Study context appears to be device engineering; population/sample details are not provided."
  ],
  "evidence_strength": "insufficient",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "wireless power transfer",
    "implantable device",
    "locomotion",
    "CMOS",
    "1.86 GHz",
    "RF power signal",
    "asynchronous PWM",
    "data rate",
    "energy efficiency",
    "magnetic field"
  ],
  "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 3136

A mm-sized wirelessly powered and remotely controlled locomotive implant. · Non-native EMF