Tailoring magnetic field gradient design to magnet cryostat geometry.

A Trakic, F Liu, H S Lopez, H Wang, S Crozier

PAPERpubmedConference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference2006engineeringEffect: unclearEvidence: Insufficient

Abstract

Eddy currents induced within a magnetic resonance imaging (MRI) cryostat bore during pulsing of gradient coils can be applied constructively together with the gradient currents that generate them, to obtain good quality gradient uniformities within a specified imaging volume over time. This can be achieved by simultaneously optimizing the spatial distribution and temporal pre-emphasis of the gradient coil current, to account for the spatial and temporal variation of the secondary magnetic fields due to the induced eddy currents. This method allows the tailored design of gradient coil/magnet configurations and consequent engineering trade-offs. To compute the transient eddy currents within a realistic cryostat vessel, a low-frequency finite-difference time-domain (FDTD) method using total-field scattered-field (TFSF) scheme has been performed and validated.

AI evidence extraction

At a glance
Study type
engineering
Effect direction
unclear
Population
—
Sample size
—
Exposure
MRI gradient coils / cryostat bore (eddy currents)
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes

Main findings

The paper describes an approach to use eddy currents induced in an MRI cryostat bore constructively with gradient currents to improve gradient uniformity within a specified imaging volume over time. It reports simultaneous optimization of gradient coil current spatial distribution and temporal pre-emphasis, and states that a low-frequency FDTD method with a TFSF scheme was performed and validated to compute transient eddy currents in a realistic cryostat vessel.

Outcomes measured

  • Gradient field uniformity over time
  • Transient eddy currents in MRI cryostat bore
  • Secondary magnetic fields due to eddy currents
  • Validation of low-frequency FDTD (TFSF) simulation method
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": null,
    "source": "MRI gradient coils / cryostat bore (eddy currents)",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "Gradient field uniformity over time",
    "Transient eddy currents in MRI cryostat bore",
    "Secondary magnetic fields due to eddy currents",
    "Validation of low-frequency FDTD (TFSF) simulation method"
  ],
  "main_findings": "The paper describes an approach to use eddy currents induced in an MRI cryostat bore constructively with gradient currents to improve gradient uniformity within a specified imaging volume over time. It reports simultaneous optimization of gradient coil current spatial distribution and temporal pre-emphasis, and states that a low-frequency FDTD method with a TFSF scheme was performed and validated to compute transient eddy currents in a realistic cryostat vessel.",
  "effect_direction": "unclear",
  "limitations": [],
  "evidence_strength": "insufficient",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "MRI",
    "gradient coils",
    "cryostat bore",
    "eddy currents",
    "gradient uniformity",
    "pre-emphasis",
    "finite-difference time-domain",
    "FDTD",
    "total-field scattered-field",
    "TFSF"
  ],
  "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 3482

Tailoring magnetic field gradient design to magnet cryostat geometry. · Non-native EMF