Electromagnetic field-based image-guided spine surgery part one: results of a cadaveric study evaluating lumbar pedicle screw placement.

H C Sagi, R Manos, R Benz, N R Ordway, P J Connolly

PAPERpubmedSpine2003OtherEffect: benefitEvidence: Low

Abstract

STUDY DESIGN: Human cadaveric. OBJECTIVES: Compare the accuracy of electromagnetic field (EMF)-based image-guided lumbar pedicle screw insertion to conventional techniques using anatomic landmarks, and fluoroscopy. BACKGROUND: Image-guided surgical systems that aid in spinal instrumentation seek to minimize radiation exposure and improve accuracy. EMF tracking-based image-guidance was developed in the hopes of eliminating line-of-sight restrictions seen with other systems. MATERIALS AND METHODS: Sixteen fresh-frozen human cadavers were randomly allocated into three groups. Pedicle screws were inserted from L1 to L5 using only anatomic landmarks in group 1, fluoroscopy in group 2, and image-guidance in group 3. Insertion and total fluoroscopic time were recorded. Anatomic dissections were performed to assess screw placement. RESULTS: Accuracy was 83%, 78%, and 95% for groups 1, 2, and 3, respectively. However, image-guided pedicle screw placement resulted in a 5% critical perforation rate whereas anatomic and fluoroscopic placement resulted in a 15% and 22% critical perforation rate, respectively. The average degree of perforation was 1.5 mm with image guidance, and 3.8 mm with fluoroscopic guidance (P < 0.05). Fluoroscopy time and insertion time per screw were not improved using image guidance. CONCLUSIONS: Our study has shown that when EMF tracking was used for image-guided lumbar pedicle screw placement, accuracy was improved and the incidence and degree of cortical perforations that may place neurovascular structures at risk was also reduced. Current system requirements for set-up and image acquisition, however, do add time to the procedure, and when factored in, do not yet result in a decrease in the use of fluoroscopy or screw insertion time.

AI evidence extraction

At a glance
Study type
Other
Effect direction
benefit
Population
Fresh-frozen human cadavers (lumbar spine instrumentation)
Sample size
16
Exposure
electromagnetic field (EMF) tracking-based image-guidance system (surgical navigation)
Evidence strength
Low
Confidence: 78% · Peer-reviewed: yes

Main findings

In 16 cadavers, EMF-based image-guided lumbar pedicle screw placement had higher accuracy (95%) than anatomic landmarks (83%) or fluoroscopy (78%). Critical perforation rates were lower with image guidance (5%) versus anatomic (15%) and fluoroscopy (22%), and average perforation was smaller than with fluoroscopy (1.5 mm vs 3.8 mm; P < 0.05). Fluoroscopy time and insertion time per screw were not improved using image guidance.

Outcomes measured

  • Pedicle screw placement accuracy
  • Critical perforation rate
  • Degree of perforation (mm)
  • Fluoroscopy time
  • Insertion time per screw

Limitations

  • Cadaveric study (not clinical outcomes in living patients)
  • Frequency/SAR and detailed EMF exposure parameters not reported
  • Fluoroscopy time and insertion time per screw not improved despite accuracy benefits
View raw extracted JSON
{
  "study_type": "other",
  "exposure": {
    "band": null,
    "source": "electromagnetic field (EMF) tracking-based image-guidance system (surgical navigation)",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": "Fresh-frozen human cadavers (lumbar spine instrumentation)",
  "sample_size": 16,
  "outcomes": [
    "Pedicle screw placement accuracy",
    "Critical perforation rate",
    "Degree of perforation (mm)",
    "Fluoroscopy time",
    "Insertion time per screw"
  ],
  "main_findings": "In 16 cadavers, EMF-based image-guided lumbar pedicle screw placement had higher accuracy (95%) than anatomic landmarks (83%) or fluoroscopy (78%). Critical perforation rates were lower with image guidance (5%) versus anatomic (15%) and fluoroscopy (22%), and average perforation was smaller than with fluoroscopy (1.5 mm vs 3.8 mm; P < 0.05). Fluoroscopy time and insertion time per screw were not improved using image guidance.",
  "effect_direction": "benefit",
  "limitations": [
    "Cadaveric study (not clinical outcomes in living patients)",
    "Frequency/SAR and detailed EMF exposure parameters not reported",
    "Fluoroscopy time and insertion time per screw not improved despite accuracy benefits"
  ],
  "evidence_strength": "low",
  "confidence": 0.78,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "electromagnetic field tracking",
    "image-guided surgery",
    "spine surgery",
    "lumbar pedicle screw",
    "fluoroscopy",
    "cadaveric study",
    "accuracy",
    "cortical perforation"
  ],
  "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 5262

Electromagnetic field-based image-guided spine surgery part one: results of a cadaveric study evaluating lumbar pedicle screw placement. · Non-native EMF