A wireless device for measuring hand-applied forces.

William Tam, Robert Allen, Y S G Hoe, I-Jean Khoo, Stanley Huang, Katherine Outland, Edith Gurewitsch

PAPERpubmedConference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference2004engineeringEffect: unclearEvidence: Insufficient

Abstract

We report on a wireless, electromyography (EMG)-based, force-measuring system developed to quantify hand-applied loads without interfering with grasping function. A portable surface EMG device detects and converts to voltage output biopotentials generated by muscle contractions in the forearm and upper arm during hand-gripping and traction activities. After amplifying and bandpass filtering, our radio frequency (RF)-based design operating at approximately 916 MHz wirelessly transmits those voltages to a data acquisition (DAQ) system up to 20 meters away. A separate calibration system is used to relate an individual user's EMG signal to known pull and clenching forces during specific applications. Real-time EMG data is processed and displayed in software developed with LabView (National Instruments, Austin, TX). Data is then converted to force data using individual calibration curves. With EMG electrodes placed over any major forearm muscle, calibration curves for seven subjects demonstrated linearity (R(2) > 0.9) and repeatability (<10% of average slope) to 110 newtons (N). Preliminary results in clinical application on newborn delivery suggest that this approach may be effective in providing an unobtrusive and accurate method of measuring hand-applied forces in applications such as rehabilitation and training.

AI evidence extraction

At a glance
Study type
engineering
Effect direction
unclear
Population
—
Sample size
7
Exposure
RF wireless device · 916 MHz
Evidence strength
Insufficient
Confidence: 74% · Peer-reviewed: yes

Main findings

An RF-based wireless EMG force-measuring system operating at approximately 916 MHz transmitted EMG-derived voltages to a DAQ system up to 20 m away. Calibration curves for seven subjects showed linearity (R^2 > 0.9) and repeatability (<10% of average slope) up to 110 N; preliminary clinical application in newborn delivery suggested the approach may be effective for unobtrusive force measurement.

Outcomes measured

  • Device performance (linearity of calibration curves)
  • Device performance (repeatability)
  • Force measurement up to 110 N
  • Wireless transmission range (up to 20 m)
  • Preliminary clinical application in newborn delivery (feasibility)

Limitations

  • No health effects were assessed
  • the RF component is part of a device design description.
  • Sample size for calibration was seven subjects.
  • Clinical application results are described as preliminary without quantitative outcomes in the abstract.
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": "RF",
    "source": "wireless device",
    "frequency_mhz": 916,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": 7,
  "outcomes": [
    "Device performance (linearity of calibration curves)",
    "Device performance (repeatability)",
    "Force measurement up to 110 N",
    "Wireless transmission range (up to 20 m)",
    "Preliminary clinical application in newborn delivery (feasibility)"
  ],
  "main_findings": "An RF-based wireless EMG force-measuring system operating at approximately 916 MHz transmitted EMG-derived voltages to a DAQ system up to 20 m away. Calibration curves for seven subjects showed linearity (R^2 > 0.9) and repeatability (<10% of average slope) up to 110 N; preliminary clinical application in newborn delivery suggested the approach may be effective for unobtrusive force measurement.",
  "effect_direction": "unclear",
  "limitations": [
    "No health effects were assessed; the RF component is part of a device design description.",
    "Sample size for calibration was seven subjects.",
    "Clinical application results are described as preliminary without quantitative outcomes in the abstract."
  ],
  "evidence_strength": "insufficient",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "wireless",
    "electromyography",
    "EMG",
    "force measurement",
    "hand-applied forces",
    "radio frequency",
    "916 MHz",
    "biopotentials",
    "calibration",
    "LabView"
  ],
  "suggested_hubs": []
}

AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract/metadata and may be incomplete or incorrect. This content is for informational purposes only and is not medical advice.

Comments

Log in to comment.

No comments yet.

NNEMF synthesis

Synthesis pending.

Original record: rfsafe.org paper 3611

A wireless device for measuring hand-applied forces. · Non-native EMF